Commissioning of the double Penning trap system MLLTRAP and first studies on mass-dependent systematic uncertainties* — •Kevin Krug1, Christine Weber1, Peter G. Thirolf1, Jerzy Szerypo1, Veli Kolhinen1,2, Eva Gartzke1, and Dietrich Habs1 — 1Fakultät für Physik, LMU München — 2Department of Physics, University of Jyväskylä
The cylindrical double Penning trap system MLLTRAP in its commissioning phase at the Maier-Leibnitz-Laboratory (MLL) Tandem accelerator in Garching is designed to perform high-accuracy mass measurements on fusion-reaction products. As the mass uncertainty is inversely proportional to the ionic charge state, the ions of interest will be charge bred prior to injection into the Penning trap system. In the future setup, both traps are foreseen to be operated as measurement traps with a relative homogeneity of the magnetic field at the trapping sites of Δ B/B ≤ 0.3 ppm. In the commissioning phase, an offline surface ionization source is used for iterative optimization of the apparatus and studies on mass-dependent systematic uncertainties. Mass measurements via the time-of-flight ion cyclotron resonance method (TOF-ICR) of reference ions with well-known masses (85Rb, 87Rb, 39K, 133Cs) were carried out to analyze mass-dependent systematic effects. Together with previous studies on the uncertainty due to magnetic-field fluctuations the current status with respect to the limits of mass accuracy of the apparatus will be presented.
* Supported by DFG under contract HA 1101/14-1 and by Maier-Leibnitz-Laboratory,Garching.