Parts | Days | Selection | Search | Downloads | Help

MS: Fachverband Massenspektrometrie

MS 2: Ionenfallen- und FT-IZR-MS, Moleküle, Cluster, Zerfälle und Reaktionen

MS 2.4: Talk

Monday, March 10, 2008, 17:30–17:45, 3E

MLLTRAP: Penning trap facility for high precision mass measurements — •Veli Kolhinen1, Michael Bussmann1, Dietrich Habs1, Jürgen Neumayr1, Ulrich Schramm2, Christian Schürmann1, Michael Sewtz1, Jerzy Szerypo1, and Peter Thirolf11Department für Physik, Ludwig-Maximilians-Universität München, Am Coulombwall 1,85748 Garching, Germany — 2Forschungszentrum Dresden-Rossendorf, 01314 Dresden, Germany

The MLLTRAP at the Maier-Leibnitz-Laboratory (Garching) will be a new Penning trap facility designed to combine several novel technologies to decelerate, charge breed, cool, bunch and purify the reaction products and perform high-accuracy nuclear and atomic mass measurements.

The double Penning trap is now in its commissioning phase at the Maier-Leibnitz-Laboratory in Garching. In the first phase we have set up the system off-line and test it with stable Cs ions. The beam has been shot through the trap system and pulsing of the walls has successfully been tested.

This contribution will present the status of the project including results from the commissioning phase and technical details as well as future plans.