Characterization of lanthanum-chloride detectors for prompt fission γ-ray measurements — •Andreas Oberstedt1, Stephan Oberstedt2, and Wouter Geerts21Örebro University, S-70182 Örebro — 2EC-JRC IRMM, B-2440 Geel

A particular challenge for the modelling of new generation reactor neutron kinetics is the calculation of the γ-heat deposition in e. g. steel and ceramics reflectors without UO2 blankets, which is required to be known with an uncertainty as low as 7.5%. A major difficulty in measuring the competition between neutron and γ-ray emission during fission fragment deexcitation is the suppression of background γ-rays induced by prompt fission neutrons in the γ-detector. A common method is to distinguish between γ-rays and neutrons by their respective different time-of-flight, which however is limited by the timing resolution of the detector (not better than 5 ns for NaI). A promising approach seems to be the use of recently developed cerium-doped lanthanum halide crystal scintillation detectors. We will present results of the characterization of coaxial 1.5″ × 1.5″ LaCl3:Ce detectors in terms of energy conversion and resolution, linearity, dynamical range, intrinsic efficiency, timing resolution and intrinsic radioactivity.